Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Installing the ESI Probe
40
LTQ XL Getting Started
Thermo Electron Corporation
Figure 14.
ESI probe, side view
4. Push the probe into the port until the guide pin meets with the probe
collar on the ion source housing.
5. Turn the probe 45 degrees clockwise and align the guide pin with the
slot in the API interlock block (you might need to pull the probe
towards you slightly to properly align the pin with the notch). Once you
have turned the probe far enough to align the pin with the alignment
notch at the rear of the port, push the probe straight in until the guide
pin stops at the bottom of the alignment notch.
6. Lock the probe in place by rotating the probe locking lever towards the
front of the housing; closing the probe locking lever towards the rear of
the ion source housing might make it difficult to unlock. You might first
need to tighten the locking lever threaded shaft by rotating it clockwise a
few turns if rotating the lever does not tighten the probe collar enough.
7. Insert the APCI vaporizer heater cable into the API interlock socket.
Auxiliary Gas
Inlet (A)
Sheath Gas
Inlet (S)
Sample
Inlet Fitting
ESI Needle High Voltage
Connector Receptacle
Sheath Liquid /
Calibrant Inlet (C)
Guide
Pin