What Types of Buffers Should I Use? What Types Should I Avoid?
10
LTQ XL Getting Started
Thermo Electron Corporation
What Types of Buffers
Should I Use? What
Types Should I Avoid?
Many LC applications use nonvolatile buffers such as phosphate and borate
buffers. It is best to avoid the use of nonvolatile buffers with the MS
detector because they can cause the following problems:
• Blocking the capillary in the probe
• Causing salt buildup on the spray head and thus compromising the
integrity of the spray
Use volatile buffers when you use the MS detector. Many volatile buffer
solutions are available that can be used instead of nonvolatile ones. Volatile
buffer solutions can include the following:
• Acetic acid
• Ammonium acetate
• Ammonium formate
• Ammonium hydroxide
• Triethylamine (TEA)
• Trifluoroacetic acid