5
Acquiring ESI Sample Data Using the Tune Plus Window
Setting Up to Acquire MS/MS Data in the Full Scan Type
90
LTQ XL Getting Started
Thermo Electron Corporation
2. Click
OK
to close the message box, and leave the Tune dialog box open.
Your Tune Plus window should now look similar to the one shown in
.
3. In the Spectrum view of Tune Plus, observe the MS/MS Full scan
spectrum of reserpine (or that of your analyte of interest).
4. When the collision energy is optimized, the Accept Optimized Value
dialog box appears. See
.
5. Click
Accept
to accept the new collision energy value, and return to
Tune Plus. The new value is displayed in the Define Scan dialog box.
6. In the Syringe Pump dialog box, select the Off option button to turn off
the syringe pump. Click
Close
to close the Syringe Pump dialog box.
7. Click
Cancel
to close the Tune dialog box.
After you optimize the relative collision energy, the LTQ XL MS detector is
ready to acquire MS/MS data on your analyte of interest.