Operation
Application Screens
The Additional Measurement tab allows the user to assign measurement inputs to
measurements two, three and four.
1.
The user may choose the display units for Measurements #2, #3 and #4.
2.
Additionally, the user may enter a level measurement Minimum Span and
Maximum Span for Measurements #2, #3 and #4.
3.
Use the fields under Number of Decimal Places to determine the number of
decimal places to display. These fields are configurable from 0 to 4.
4.
In each instance where the unit inputs are set to Not Used, then Measurement #1
will be used on the Measurement Data tab.
General Setup
The General setup screen allows the user to read and modify parameters associated
with the system. The LevelPRO detector will be read and will write any of these
parameters as requested by the user.
To access the information on the Application General Setup tabs using only the
remote transmitter keypad, see
Figure 4-26.
Application Screen, General Setup Tab
1.
Select a source type from the Source Type (Hi Voltage Control) dropdown.
•
Cs137 (Cesium 137)
•
Co60 (Cobalt 60)
•
Am241 (Americium 241 Beryllium)
•
Other
2.
The user has the option of entering the requested information into the following
textboxes:
•
Radioactive Source Serial Number
•
Radioactive Source Tag Number
•
Radioactive Source Assay Date (MM/YY)
4
-
44
LevelPRO User Manual
Thermo Scientific