ISQ EC and ISQ EM Mass Spectrometer Preinstallation Requirements Guide
Page 31
Figure 1: Location of the feet on the bottom of the ISQ EC or ISQ EM instrument
TIP
—Thermo Fisher Scientific requires that all six of the instrument's
feet be positioned on a bench. Overhang of the front, the back, or both
ends of the instrument off the bench front or back is acceptable.
The foreline hose exits at the rear of the instrument and is connected to
the foreline pump. The hose can be moved to exit to the right, the left,
or back.
The foreline pump is placed on the floor below or next to the system.
The foreline hose is 250 cm long measured from the back of the MS. It
cannot be extended because that will adversely affect the pressure in
the MS. See
and
for front and back views of the ISQ EC
or ISQ EM system with all the hoses and gas lines attached.