1
Performing Routine Maintenance
Cleaning the Ion Source Cartridge
52
ISQ 7000 Hardware Manual
Thermo Scientific
5. Press the blue
Evacuate
button on the front of the instrument. When the Evacuate light
flashes green for approximately 20-30 seconds and turns off, the
Ready to Open
light
turns solid green. At that point, the air has been evacuated from the instrument and it is
safe to open the vacuum interlock valve.
6. Loosen and twist the top part of the vacuum interlock handle to allow the entire handle
to move up and down. This handle controls the interior ball valve, which seals the
vacuum chamber and protects it from the outside pressure.
Figure 35.
Evacuating the Instrument
Evacuate
Button
Vacuum
Interlock
Handle
Содержание ISQ 7000
Страница 1: ...ISQ 7000 Mass Spectrometers Hardware Manual 1R120617 0002 Revision C December 2018 ...
Страница 8: ......
Страница 174: ...2 Troubleshooting Contacting Technical Support 146 ISQ 7000 Hardware Manual Thermo Scientific ...
Страница 400: ...3 Advanced Maintenance Replacing Fans and Filters 372 ISQ 7000 Hardware Manual Thermo Scientific ...