5. Remove the vacuum interlock knob and then use a T10 Torxhead screwdriver to remove
the five screws around the knob.
Note
As you remove the screws around the vacuum interlock knob, watch for a small
spring and safety interlock (location shown below) that will fall out as your remove the
knob. Be sure to catch them and set them aside.
Figure 303.
Removing the Vacuum Interlock Knob
Screws
Vacuum
Interlock Shield
Spring and safety interlock
are behind this screw