3
Advanced Maintenance
Replacing Components of the Manifold
240
ISQ 7000 Hardware Manual
Thermo Scientific
7. Pinch together the two springs on the left side the vacuum interlock shield to remove the
shield and door.
8. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
remove the three screws connecting the hinge to the chassis.
Figure 245.
Removing the Vacuum Interlock Shield and Manifold Door
Figure 246.
Replacing the Manifold Door Hinge
Springs
Vacuum
Interlock
Shield
Screws
Hinge