Replacing Components of the Vacuum Interlock
228
ISQ 7000 Hardware Manual
Thermo Scientific
15. Flip the block over in your hand so that sealing ball and outer ball seal will fall out. Set
them aside or replace them, if necessary.
16. Pry the thin metal clip out of the groove. It holds all of the other components inside the
valve body.
CAUTION INSTRUMENT DAMAGE:
The sealing ball must not be scratched, so be very
careful when handling it and keep it on a clean, lint-free surface.
ATTENTION DOMMAGES À L’INSTRUMENT
: la bille d’étanchéité ne doit pas être rayée,
soyez donc très prudent lorsque vous la manipulez et maintenez-la sur une surface propre et
non pelucheuse.
Figure 227.
Replacing the Sealing Ball and Outer Ball Seal
Figure 228.
Replacing the Metal Clip
Sealing Ball
Outer Ball Seal
Clip