Advanced Maintenance
Replacing Components of the Vacuum Interlock
224
ISQ 7000 Hardware Manual
Thermo Scientific
3. Disconnect the vacuum interlock cable from the connector.
4. Pinch together the two springs on the left side the vacuum interlock shield and remove it.
5. Lift the vacuum interlock shield off the instrument. The notch at the bottom of the track
will go around the handle.
Figure 220.
Disconnecting the Vacuum Interlock Cable
Figure 221.
Removing the Vacuum Interlock Shield
Vacuum
Interlock
Cable
Connector
Springs