Troubleshooting
Investigating Power Supply Issues
136
ISQ 7000 Hardware Manual
Thermo Scientific
Investigating Power Supply Issues
When you run Power Supply Check diagnostics on the ISQ 7000 system, the Xcalibur
software verifies that all power supply voltages are within the acceptable ranges. Your
instrument or software may indicate the following types of issues.
The ISQ 7000 instrument will not power-on.
The ISQ 7000 instrument powers on temporarily and trips the circuit breaker.
±Q1 DC supply is outside the acceptable range.
WARNING ELECTRICAL SHOCK HAZARD:
When troubleshooting any issue that requires
removing a cover on the ISQ 7000 mass spectrometer, you should power-off and vent the
instrument to avoid physical harm.
AVERTISSEMENT RISQUE D’ÉLECTROCUTION
: lorsque vous cherchez à résoudre un
problème nécessitant d’enlever le couvercle du spectromètre de masse ISQ 7000, mettez
hors tension et ventilez l’instrument pour éviter de vous blesser.
Possible Remedies
Make sure the ISQ 7000 instrument’s power cable is properly connected to the instrument and electrical outlet.
Verify that the electrical outlet is functioning properly.
Replace the ISQ 7000 instrument’s power supply system. See
for details.
Possible Remedies
Make sure the ISQ 7000 instrument’s power cable is not damaged.
Replace the ISQ 7000 instrument’s power supply system. See
for details.
Check the foreline pump voltage setting and replace the pump, if necessary.
Possible Remedies
Make sure the Rod Driver board is properly connected and use the Power Status tab on the ISQ 7000 Dashboard to ensure the readback
r48 V DC is present.
Note: Some power supplies remain off until the vacuum is OK and the turbo pump speed is above 80%. This includes the +- 75 V, +- 175 V,
and +- Q1 DC supplies.