Performing Routine Maintenance
Maintaining the Calibration Gas Module
Thermo Scientific
ISQ 7000 Hardware Manual
89
Refilling the Calibrant Reservoir
To refill the calibrant reservoir
1. Open the front door of the ISQ 7000 instrument.
2. Twist the calibrant reservoir cover counter-clockwise and remove it from the calibration
gas module.
3. Fill a syringe with 100 μL of Perfluorotributylamine/PFTBA/FC 43 and insert it into the
calibrant reservoir.
Figure 87.
Removing the Calibrant Reservoir Cover
Figure 88.
Removing the Calibrant Reservoir from the Cover
Calibrant
Reservoir
Cover
Calibration
Gas Module
Calibrant
Reservoir
Cover
Calibrant
Reservoir