Confirm that consumables are correctly installed
• Confirm that each cartridge is at the correct location and in the correct orientation.
• Press down on all cartridges to confirm that they are firmly pressed into place.
• Confirm that all tubes in the Ion S5
™
ExT Chef Reagents cartridge, including the tube of NaOH in
Position C, are uncapped and firmly pressed into place.
• Confirm that the centrifuge lids are installed correctly so that the port is oriented toward the rear of
the instrument.
• Confirm that the tube and chip buckets are seated securely in the rotor arms of the Chip-loading
and Recovery centrifuges, and that the consumables they contain are correctly installed.
CAUTION!
To ensure correct and safe instrument operation, you must confirm that all consumables
are installed correctly to the deck before you start a run. The Ion Chef
™
Instrument does not verify all
aspects of the consumable setup before beginning each run.