Load the Ion Chef
™
System
IMPORTANT!
·
Rated centrifuge speeds are intended for operation only with the provided buckets and approved
consumable chips, tubes, and sample preparation reagents.
·
The Chip-loading centrifuge is rated to operate at the listed rotational frequencies with the chip
buckets, chip, adapter, and chip balance. The centrifuge must be load-balanced. Proper care must
be taken to load the bucket properly. If excessive vibrations arise, check to ensure that items are
installed properly and rotors are load-balanced.
·
Use only the materials supplied in the Ion 520
™
& Ion 530
™
ExT Kit – Chef to run the centrifuges at
the rated speeds. Do not remove or change the rotors. Inspect the buckets before each use to assure
normal operation.
·
Ensure that the instrument is powered on and has been cleaned following the last use.
·
Ensure that all components are clean and dry before loading them onto the Ion Chef
™
Instrument.
·
Ensure that the Reagents and Solutions station compartments are dry and free of condensate before
loading components.
Follow the procedure described in the following sections to load the Ion Chef
™
Instrument.
A
B
C
D
E
F
G
H
1
2
3
4
5
6
7
8
9
10
11 12
530
DABD02226
Ion
Che
f™
S5 Series
Chip Balance
5
9
3
1
2
4
6
7
8
Figure 1 A schematic of a loaded Ion Chef
™
Instrument.
1
New Ion S5
™
ExT Tip Cartridge
2
Frame Seal v2 (inserted)
3
Empty tip rack (move from new Tip cartridge position)
4
PCR Plate
5
Ion S5
™
ExT Chef Reagents cartridge
6
Ion S5
™
Chef Solutions cartridge
7
Recovery centrifuges: Recovery Tubes and Recovery
Station Disposable Lid v2
8
Enrichment Cartridge v2
9
Chip-loading centrifuge: Chip Adapter/Chip assembly
and Chip Balance
Chapter 4
Start the Ion Chef
™
run
Load the Ion Chef
™
System
4
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide
27