Change light cube
The Countess
™
II FL Automated Cell Counter can accommodate up to two EVOS
™
light cubes. Each user-interchangable, auto-configured EVOS
™
light cube contains an
LED, collimating optics, and filters for fluorescence applications. EVOS
™
light cubes
do not come standard with the device and must be purchased separately (“EVOS
light cubes“ on page 64). To install or change a light cube:
1.
Press the Instrument Settings button on the Home page (“Turn ON the
instrument“ on page 13) to open the Instrument Settings screen.
2.
Press Change Light Cube. The instrument positions the light cube tray to enable
light cube installation.
3.
When prompted, power off the Countess
™
II FL Automated Cell Counter using
the power switch on the back of the instrument (“Instrument exterior
4.
Unplug the power cord from the Countess
™
II FL Automated Cell Counter.
5.
Unlatch the back panel with the two captive ¼-turn fasteners (indicated by red
arrows) that secure the back panel on the rear of the Countess
™
II FL Automated
Cell Counter and remove the back panel.
6.
Place the light cube into one of the empty slots in the back of the device.
Install or change
EVOS
™
light cube
Chapter 8
Instrument settings
Change light cube
8
54
Countess
™
II FL automated cell counter User Guide