Getting started
Installation
• Place the instrument on a level surface away from vibrations emanating from
other pieces of equipment.
• Allow at least 5 cm (2 in) free space at the back of the instrument to allow for
proper ventilation and prevent overheating of electronic components.
• Set up the instrument away from direct light sources, such as windows. Ambient
room lighting can enter the imaging path and affect the image quality.
• Operating temperature range: 4°–32
℃
(40°–90°F).
• Relative humidity range: <80%.
IMPORTANT!
Do not position the instrument so that it is difficult to turn off the main
power switch located on the back of the instrument (“Instrument exterior
components“ on page 11).
In case of an instrument malfunction, turn the main power switch to the OFF position
and disconnect the instrument from the wall outlet.
1.
Unpack the instrument and place the instrument on a flat, level, dry surface.
2.
Remove the thin plastic protector film from the touch-screen display.
3.
Plug one end of the power cord appropriate for your region into the instrument.
4.
Plug the power cord into the electrical outlet. Be sure to use only the power cord
supplied with your instrument. Powering the instrument with an unapproved
power cord may damage the instrument.