Preparing the Work Space
5-6 Heratherm
Thermo Electron LED GmbH
Figure 5-8
Removing the Lateral Air Baffle
3. For the lateral air baffles to install correctly, the two retaining springs [1] must be facing
upwards. Place the lateral air baffle on the lower embossings and tilt it upwards against
the side wall of the work space.
4. Clamp the two retaining springs [1] into the upper embossings.
5. Replace the bottom panel into the embossings and secure it by fastening the two screws
at [1].
Installing the Shelf Support Brackets
1. Insert the shelf support brackets [3] into the perforations [1] of the support rail and air
baffle and tilt them downwards.
2. Make sure that the two vertical elements [2] of the shelf support bracket butt against the
support rail and air baffle.
Horizontal position of support rails
Please note the alternating sequence of round and oblong holes. Be
sure to use the correct pair of opposite holes, so the support rails are
truly horizontal.
Содержание Heratherm OGS 100
Страница 1: ......
Страница 6: ...Contents 4 Heratherm Thermo Scientific ...
Страница 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Страница 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Страница 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Страница 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Страница 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Страница 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Страница 84: ...Device Log 15 2 Heratherm Thermo Scientific ...
Страница 87: ...Contact Thermo Scientific Heratherm 16 3 ...
Страница 88: ...Contact 16 4 Heratherm Thermo Scientific ...