1
Spectrophotometer Basics
Z-dimensions
6
GENESYS 10S Vis User Guide
Thermo Scientific
Other Guidelines
Position cuvettes and test tubes so that the clear sides face the light beam, one clear side facing
the front of the instrument and the other facing the back.
When using small aperture (small volume) cells:
• Always used cells with black masking
• Use the same cell (or cuvette) for your blank and your samples
Z-dimensions
The figure below illustrates the position of the light beam in the instrument.
Beam size specifications are shown below.
• Distance from bottom of cuvette to center of beam (Z-dimension): 8.5 mm
• Beam dimensions: 2 mm (wide) by 7 mm (tall)
Note
The pathlength of test tubes is not as well defined as that of square cuvettes.
Note
Always place test tubes in the instrument in exactly the same orientation in the light
beam. An alignment mark on the test tube helps you orient the test tubes consistently and
correctly.
Содержание GENESYS 10S Vis
Страница 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Страница 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 22: ...2 Setting Up the Instrument Setting Up the Internal Printer 14 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 56: ...11 Abs and T Measurements Advanced A T C Taking Measurements 48 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 64: ...13 Concentration Measurements Advanced A T C Measuring Samples 56 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 106: ...20 Kinetics Rescaling and Recalculating Tabular Kinetics Results 98 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...