Chapter 2
The FOCUS GC User Interface
Run-Log
64
1. Parameters available only with the TCD detector
To program a Run Time Event, refer to Chapter 11
.
Run-Log
This function displays the run log, which records errors or method changes that
happen in run time. It display the time and describes any deviation that occur.
This information is used to meet good laboratory practice (GLP) standards.
For example, if you interrupt the run for any reason, the run log will record the
time the run stopped and a description of the event. The Run-log is cleared and
reset at the beginning of the next run.
Run log entries are shown pressing the
STATUS
key.
To see the details Run-Log entries, in Main Menu move the cursor until the
Others
page is selected, then press
ENTER
. Scroll to
Run-log
then press
ENTER
An example of Run-log is shown below.
Setpoint
From 0 to 3 for Det range
On/Off for EV
Select the correct setpoint according to the event
you have set in the previous line. This line doesn’t
appear for Det autozero.
Next event
---
Add a new event
OTHERS
Run-log
Table 2-14. Edit Time Event Menu (Continued)
Parameter
Range
Description
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...