Chapter 15
Maintenance and Troubleshooting
Maintaining a FID
292
OPERATING SEQUENCE
Replace the FID Ignition Assembly
Material needed
•
Screwdriver
•
Connector Holder Assembly
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
When handling organics solvents you must take precautions to avoid health
hazards.
1. In
DETECTOR
menu, scroll to
Flame
and set OFF.
2. Scroll to
Temp
and set
OFF.
3. Scroll to
H2
and
Air
and set OFF.
4. Disconnect the signal and ignition polarization cables from the detector.
See Figure 15-19.
5. Loosen the fixing screw on the front of the detector cell and remove it.
See Figure 15-19.
DETECTOR
Flame Off
Temp 25 Off
Signal pA (0.0)
Ing. Thresh. 2.0
Flameout retry Off
H2 Off
Air Off
Make-up Off
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...