Chapter 12
SOP number: P0407/05/E - 31 May 2006
220
OPERATING PROCEDURE
FID-S/SL Checkout in Splitless Mode
Before starting, verify the configuration of the carrier and set the checkout
required parameters.
Verify Carrier Configuration
1. In Main Menu move the cursor until
Others
is selected, then press
ENTER
.
Scroll to
Configuration
, then press
ENTER
to open Configuration menu
.
Scroll to
Configure carrier
submenu then press
ENTER
.
Select the carrier gas required by using the
INC
and
DEC
keys.
.
2. Press
HOME
three times to return Main Menu.
Set Checkout Parameters
3. In Main Menu move the cursor until
Carrier
is selected. Press
ENTER
to
open
CARRIER
menu.
FOCUS GC
Others
OTHERS
Configuration
CONFIGURATION
Carrier
CONFIGURE CARRIER
Gas type
Helium
FOCUS GC
Carrier
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...