Chapter 9
Thermal Conductivity Detector (TCD)
TCD Operating Modes
176
temperature limit in the TCD detector menu (Table 9-4). If this value is reached,
the system will automatically switch to the constant temperature mode.
The CV mode extends the linearity range to higher values, but with a negative
impact on sensitivity.
NOTE
Automatic Switching to Constant Current Mode
Every time the set values of filaments voltage, block temperature and filaments
temperature cause the filament current to reach the maximum value of 125 mA, the system
will automatically switch to the constant current mode and the filaments cannot be heated
more than the correspondent temperature.
This mode has good sensitivity and a linearity comparable to that obtained with the CV
mode. However, the high filament temperatures can potentially shorten the filament life.
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...