Flame Ionization Detector (FID)
Detector Signal Parameters
166
Detector Signal Parameters
The detector
SIGNAL
menu contains the parameters that control the detector
signal. As compounds elute from the column and enter the detector, an electrical
signal is generated. The size of the signal is related to the amount of the
corresponding compounds. The detector’s electronics process the signal and send
it to a recording device. The plot of the signal size versus the time results in the
chromatogram.
In Main Menu move the cursor until
Signal
is selected. Press
ENTER
to display
the
SIGNAL
menu.
SIGNAL
Output (1000000)
Offset 100000
Run autozero On
Range = 10^ (0..3) 0
Table 8-2. FID Signal Parameters
Parameter
Range
Description
SIGNAL
This line is the title bar
Output
Not editable
This is the actual electrometer output signal expressed in
gV. The
Autozero
function forces this value to 1000
corresponding to the
zero level of the baseline on a
recording device. You cannot enter a setpoint here.
Offset
0 to 100000
This is a value in counts that may be subtracted from the
Output signal to adjust the baseline level. This
parameter may be manually or automatically set using
the
Auto zero
function. The range of the suppression
is variable and related to the output signal.
Run autozero
On - Off
This function forces the output signal to 1000 (zeroing).
Set On to zero the detector signal.
Range
10 0 - 10 3
1, 10, 100, 1000 nA
This parameter sets the electrometer amplifier input
range. 10 0 are the most sensitive.
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...