Eutech DO 1730 Dissolved Oxygen Meter     23 

3.2.16 Audio Mode 

Set the audible beep on or off. The audible beep is used 
when an alarm is triggered. 

 

Press the 

 or 

 key to scroll through the Setup 

Menu list until the Audio Mode item is shown. 

 

 

Press the 

 or 

 key to set the audio mode. 

 

 

Press the “Enter” key to save the change. 

 

3.2.17 Sleep Mode 

Set the sleep mode on or off. When the meter sleep mode 
is on, the meter will enter sleep mode when no keys are 
pressed for 20 minutes. Once the meter is in sleep mode, 
press the “Power” key to resume using the meter. Make 
sure sleep mode is off when taking timed measurements. 

 

Press the 

 or 

 key to scroll through the Setup 

Menu list until the Sleep Mode item is shown. 

 

 

Press the 

 or 

 key to set the sleep mode. 

 

 

Press the “Enter” key to save the change. 

 

Содержание Eutech DO 1730

Страница 1: ...Operations Manual Electrochemistry Meters 68X002228 Revision 01 August 2022 Eutech DO 1730 Dissolved Oxygen Meter User Manual ...

Страница 2: ......

Страница 3: ... 9 2 3 1 Setup Menu Scrolling 9 2 3 2 Setup Menu Value Changes 9 2 3 3 Mode Shortcut Key 9 2 3 4 Date and Time View Shortcut Key 9 2 4 Display Overview 10 2 5 Meter Connections 11 2 6 Electrode Information 11 2 7 Meter Startup Sequence 11 Chapter 3 12 Setup Menu 12 3 1 Setup Menu Overview 12 3 2 Setup Menu Items 13 3 2 1 View Logs 13 3 2 1a Viewing the Data Log 13 3 2 1b Viewing the Calibration Lo...

Страница 4: ...16 Audio Mode 23 3 2 17 Sleep Mode 23 3 2 18 Clear Data 24 3 2 19 Factory Reset 24 Chapter 4 25 Calibration 25 4 1 DO Calibration 25 4 1 1 DO Calibration Overview 25 4 1 2 DO Cal Procedure for 100 Saturation 25 4 1 3 DO Cal Procedure for 0 Saturation 26 4 1 4 DO Cal Procedure for Concentration 26 4 1 5 DO Calibration Example 27 4 2 Temperature Calibration 28 Chapter 5 29 Measurements 29 5 1 Read T...

Страница 5: ...ter Dimensions 35 7 3 Warranty 35 7 3 1 Return of Items 35 Chapter 8 36 Regulatory Compliance 36 8 1 European Union 36 8 2 Product Safety 36 8 3 Electromagnetic Compatibility 36 8 3 1 FCC Statement USA 36 8 4 Canadian ISED IC Notice 36 8 5 Environmental Compliance 36 8 5 1 REACH Europe 36 8 5 2 RoHS Europe 36 8 6 WEEE Compliance 37 Chapter 9 38 Ordering Info 38 9 1 Meter Ordering Information 38 9 ...

Страница 6: ...ould result in minor to moderate injury or damage to the equipment Note This symbol in the context of a Note indicates to follow the manufacturer specified instructions notes and requirements set out in the instruction manuals CAUTION Before using or maintaining this product please be sure to read the manual carefully Failure to follow these instructions may cause the product to malfunction CAUTIO...

Страница 7: ...on a clean dry surface Identify the side of the meter that the holder will be installed on and remove the screw between the circles on that side of the meter Align the base plate of the holder with the circles on the meter Replace the screw from step 3 to attach the base plate to the meter Turn the meter over with meter display facing up Insert the stand with electrode holder into the metal post o...

Страница 8: ... Unpack the power supply provided with the meter Select the appropriate wall plug plate for the power outlet that will be used Slide the appropriate wall plug plate into the groove on the back of the power adapter Connect the assembled power adapter to the meter input Locate the two slots on the meter power input Align the two tabs on the power adapter connector with the two slots on the meter pow...

Страница 9: ...tion mode Press to power on the meter When the meter is on press to turn the backlight off or on When the meter is on press and hold for three seconds to power off the meter 2 3 1 Setup Menu Scrolling To scroll through the Setup Menu list Press the Menu key to enter the Setup Menu The first Setup Menu item will always be show Press the key to scroll to the second item or press the key to scroll to...

Страница 10: ... data log or calibration log Log Saved Icon Indicates meter is actively saving data to the data log or calibration log Continuous Icon Indicates the current meter read type is Continuous and measurement values are continuously updated on the display Salinity Icon Indicates the salinity value of the sample is being entered Stopwatch Icon Indicates stability criteria for current measurement is stabi...

Страница 11: ...electrode instructions for specific care and maintenance procedures The following are general recommendations for electrode preparation Remove the packaging from the electrode and save for storage Unscrew the membrane cap from the electrode Rinse the inside and outside of the cap with distilled or deionized water and blot dry Rinse the anode cathode of the electrode with distilled or deionized wat...

Страница 12: ... device 4 Temperature Calibration or Manual Temperature Input When DO electrode with built in ATC sensor is connected use the Temp Calibration menu to perform a temperature offset calibration When DO electrode without built in ATC sensor is connected use the Manual Temp Input menu to enter the sample temperature value 5 Measure Mode Set the main measure mode to dissolved oxygen as saturation mg L ...

Страница 13: ...ress key Press key Press the Enter key to view the display log 3 2 1a Viewing the Data Log Press the Enter key to view the display log For the data log press the or key to scroll through the data log points Scroll through data log points when viewing the date time screens or the measurement data screens Press key For the data log press the or key to change the view from the date time screen to the...

Страница 14: ... change the view from the date time screen to the calibration data screen Press or key For the calibration log press the or key to scroll through the calibration log data Press key Press key Press key If desired press the Log Export key to export the individual calibration log being viewed Press the Menu key to return to the main Setup Menu list ...

Страница 15: ...off Press the Enter key to save the change 3 2 3 Export All Logs Send all saved data logs and calibration logs to the selected export device Note that this menu will not be shown if the Log Export Type menu is set to off Press the or key to scroll through the Setup Menu list until the Export All Logs item is shown Press the Enter key to export all saved logs to the selected export device a The dis...

Страница 16: ...emperature offset value will be shown in the 2nd field a Press the key to increase the value by 0 1 Press and hold the key to quickly increase b Press the key to decrease the value by 0 1 Press and hold the key to quickly decrease Press the Enter key to save the change 3 2 4b Manual Temperature Input When DO electrode without built in ATC sensor is connected use the Manual Temperature Input menu t...

Страница 17: ...nter key to save the change 3 2 6 DO Resolution Set displayed dissolved oxygen resolution to 0 1 or 1 saturation 0 01 or 0 1 mg L or ppm The dissolved oxygen units shown in this Setup Menu item are related to the previous setting selection in the Measurement Mode item For this example dissolved oxygen in concentration ppm is shown Press the or key to scroll through the Setup Menu list until the DO...

Страница 18: ...is used for barometric pressure corrected dissolved oxygen measurements The barometric pressure units shown in this Setup Menu item are related to the previous setting selection in the Barometric Pressure Units item For this example barometric pressure in mmHg is shown Press the or key to scroll through the Setup Menu list until the Barometric Pressure Value item is shown Press the or key to adjus...

Страница 19: ...re continuously updated on the display and the stopwatch stabilizing or checkmark stable icon indicates the measurement stability status Press the Log Export key to save a measurement to the data log and export Auto Read In the Auto Read measure mode press the Measure key to start a measurement When the measurement is stable the checkmark stable icon is shown and the measurement is locked on the d...

Страница 20: ...ime value by one second Press and hold the key to quickly decrease Press the Enter key to save the change 3 2 12 Calibration Due Alarm Set the calibration due alarm interval from 0 hours off to 168 hours An alarm is triggered if a calibration is not performed within the specified time interval Press the or key to scroll through the Setup Menu list until the Calibration Due Alarm item is shown Pres...

Страница 21: ... scroll through the Setup Menu list until the Set Date Value item is shown The month value will flash Press the or key to adjust the month a Press the key to increase by one month Press and hold the key to quickly increase b Press the key to decrease by one month Press and hold the key to quickly decrease Press the Enter key to save the change The day value will flash Press the or key to adjust th...

Страница 22: ...the Enter key to save the change The hour value will flash Press the or key to adjust the hours a Press the key to increase by one hour Press and hold the key to quickly increase b Press the key to decrease by one hour Press and hold the key to quickly decrease Press the Enter key to save the change The minutes value will flash Press the or key to adjust the minutes a Press the key to increase by ...

Страница 23: ...nter key to save the change 3 2 17 Sleep Mode Set the sleep mode on or off When the meter sleep mode is on the meter will enter sleep mode when no keys are pressed for 20 minutes Once the meter is in sleep mode press the Power key to resume using the meter Make sure sleep mode is off when taking timed measurements Press the or key to scroll through the Setup Menu list until the Sleep Mode item is ...

Страница 24: ... erased The meter will confirm when the selected log is fully erased The meter will return to the measurement screen 3 2 19 Factory Reset Erase all data logs calibration logs and settings and return the meter to its factory default state Press the or key to scroll through the Setup Menu list until the Factory Reset item is shown Press the Enter key to confirm the selection Press the Enter key agai...

Страница 25: ... If the barometric pressure setting is changed the meter will automatically adjust to a new percent saturation calibration value corrected for the adjusted barometric pressure The following table provides calibration values for two barometric pressure values Note that the percent saturation value 92 1 decreases with a lower barometric pressure value Saturation of factory default value Calibration ...

Страница 26: ...ay the calculated offset value and then proceed to the main measure mode Note Press the Measure key at any time to abort the calibration and return to the main measure mode 4 1 4 DO Cal Procedure for Concentration For a dissolved oxygen as concentration calibration a minimum value of 2 0 mg L ppm must be entered and the entered concentration value must be within 40 of the factory default concentra...

Страница 27: ...ce into the prepared calibration chamber Wait for the DO value to stabilize a When the reading is unstable the stopwatch icon is shown and the reading flashes b When the reading is stable the checkmark icon is shown and the reading is solid Once the reading is stable press the Enter key The meter will save the calibration The calculated slope value is shown and then the Calibration Done screen is ...

Страница 28: ...s the or key to scroll through the Setup Menu list until the Temp Calibration item is shown The temperature value will blink indicating the reading from the ATC probe is being measured Once the reading is stable the value will stop blinking and the checkmark icon will be shown Press the or key to adjust the temperature value The temperature offset value will be shown in the 2nd field a Press the k...

Страница 29: ...issolved oxygen as percent saturation dissolved oxygen as concentration in mg L and dissolved oxygen as concentration in ppm In the main measure mode press and hold long press the or key for approximately three seconds to change the main measure mode to dissolved oxygen as saturation concentration in mg L or concentration in ppm Dissolved oxygen levels vary with temperature barometric pressure and...

Страница 30: ...g viewed 5 4 Data Exporting Export data from the meter to a computer or printer Use the Log Export Type item in the meter Setup Menu to set the export type as computer or printer If computer is selected logs are exported in CSV format If printer is selected logs are exported in list format Meter serial communication protocol Computer Setting Printer Setting Baud Rate 9600 bps Baud Rate 9600 bps Da...

Страница 31: ...ment immediately GETMEAS Data Count CR Prints the current measurement for a set number of times Data Count 1 2 3 etc GETMEAS 2 GETCAL CR Prints all current calibration data If no calibration is saved returns to receive next command GETCAL MODE CR Prints the calibration data for specific mode MODE PERC or MGL or PPM GETCAL PERC GETLOG CR Prints all logged measurement data Output format is based on ...

Страница 32: ...mofisher com orionsoftware Unzip extract the Eutech 1700 series meter software files to the computer s desktop Open the folder containing the unzipped extracted software files and double click the setup exe file to install the setup program onto the computer Follow all prompts to install the setup program onto the computer If any security warning messages appear select the run software option Afte...

Страница 33: ...he sample solution with the electrode membrane and temperature sensor submerged below the top of the sample solution o Verify cables do not have any damage o Perform a factor reset on the meter 6 2 Measurement Troubleshooting Reading is unstable slow to stabilize Clean electrode to remove build up or contaminants Replace DO electrode sensing membrane and electrolyte solution Ensure DO electrode is...

Страница 34: ...Yes Time format 12 hour AM PM Date format Selectable DD MM YYYY or MM DD YYYY Read types Continuous Auto Read Timed Timed interval 5 seconds to 60 minutes Data log 500 data sets with time and date Calibration log Active DO as saturation DO as concentration temperature with time and date Calibration alarm Yes 1 to 168 hours or off Sleep mode Yes optional Self test Automatic with startup Memory Non ...

Страница 35: ...be repaired on a charged basis The warranty on your meter shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products 7 3 1 Return of Items Authorization must be obtained from our Technical Support Team or authorized distributor before returning items for any reason ...

Страница 36: ...ironment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense 8 4 Canadian ISED ...

Страница 37: ...tsorgt werden Mehr Informationen über die Einhaltung dieser Anweisungen durch Thermo Scientific dieVerwerter und Hinweise die Ihnen nützlich sein können die Thermo Fisher Scientific Produkte zu identizfizieren die unter diese RoHS Anweisungfallen finden Sie unter www thermofisher com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2012 19 EU in merito ai Rifiuti ...

Страница 38: ...0 60Hz power adapter EB1700PWR Eutech 1700 series USB computer cable EB1700USB Eutech 1700 series meter dust cover EB1700CVR Eutech compact thermal printer with RS232 cable and paper 100 240V 01X594301 Eutech RS232 cable for thermal printer 30X427301 Eutech paper for thermal printer 2 pack 01X594401 9 2 Electrode and Solution Ordering Info Description Catalog Number Eutech galvanic DO ATC electrod...

Страница 39: ..._________________________________________________________________________________ _____________________________________________________________________________________ _____________________________________________________________________________________ _____________________________________________________________________________________ ____________________________________________________________...

Страница 40: ... of a wide variety of samples Find out more at thermofisher com electrochemistry Contact us today at customerservice sg wai thermofisher com This product is intended for General Laboratory Use It is the customer s responsibility to ensure that the performance of the product is suitable for customers specific use or application 2022 Thermo Fisher Scientific Inc All rights reserved All trademarks ar...

Отзывы: