Installing Flash IRMS Elemental Analyzer
Installing the Reactors into the Furnaces
56
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Installing the Reactors into the Furnaces
Before installing the reactors for pyrolysis and combustion the following preliminary operation must be
carried out.
1. Make sure that the furnaces are at room temperature.
2. Open the furnaces compartment by lifting the cover and removing the protection plate.
See
.
Figure 50.
Access to the Furnace Compartment
3. If present, remove the protection cap from the coupling union of each channel.
a.
By using a 32-mm wrench, unscrew the reactor retaining nut. Remove the protection cap, the
washer, and the conical O-ring as shown in
.
Note
In the following figures, the connections of the tubings are not visible for convenience.
Furnace Compartment Cover
Right Furnace
Left Furnace
Protection Plate
Furnace Compartment Protection Plate