Getting Familiar with your EA IsoLink IRMS System for CNSOH
Use of Gases
8
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Nominal Pressure
The nominal pressure of the Flash IRMS EA gas supplies are:
• 250-300 kPa (2.5-3 bar) for He
• 250-300 kPa (2.5-3) for O
2
Precaution for Carbon Monoxide
Install an exhaust tube on top of your ConFlo II/III/IV, as shown in of the example of
remove the toxic carbon monoxide (CO) from inside the ConFlo II/III/IV out of your working area.
Figure 5.
Exhaust Tube Connection
Precaution for Sulfur Dioxide
Note
For the nominal pressure of the gases used by ConFlo II/III/IV and IRMS Delta models
(
TF Bremen manufacturing
), refer to the relevant manuals.
Note
The pressure must be adjusted to about
4
kPa
through the reducing valves mounted at the
cylinder.
ATTENTION
Carbon Monoxide is toxic!
When working with this gas good ventilation is essential. Otherwise, the gas can be hazardous to
your health! The use of a CO detector with an alarm is strongly recommended.
Outside of
Working Area
Working Area
Fan
Exhaust tube
ConFlo II/III/IV
ATTENTION
Sulfur Dioxide is toxic!
When working with this gas good ventilation is essential. Otherwise, the gas can be hazardous to
your health!