Troubleshooting
EFC-t Module
140
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
EFC-t Module
The failures that may be generated on the EFCt Module are connected to the breakage, or to the
malfunctioning of solenoid valves and flow sensors.
Refer to the
to find the component responsible of the EFC-t module malfunctioning and to
solve the relevant problem.
Right furnace over limit
(Safety Cut Off message)
The right thermocouple is
damaged or interrupted.
Verify the continuity of the
thermocouple and replace it.
Thermal pre-protection
(Safety Cut Off message)
Anomalous temperature inside the
transformers compartment.
Check the functionality of the fans.
Replace the AS 1112 board.
Oxygen pressure too low
(NO Safety Cut Off message)
If the oxygen pressure is too low
(<150 kPa) the flow of oxygen does
not reach the set point.
Verify that the pressure at the
manometer is at second stage
higher or equal to 350 kPa.
Table 26.
Error Messages (Sheet 2 of 2)
Message
Description
What to do
CAUTION
The error message is generally due to the specific cause indicated. Sometimes, it may
generated by different electric factors or caused by failures not depending on the system. In this case
contact the Technical Service.
Table 27.
EFC-t module troubleshooting
Failure
Defective Component
Remedy
Oxygen does not flow to the point 2
of the autosampler
EV1
Check voltage supply
Replace the solenoid valve
EV2
Check voltage supply
Replace the solenoid valve
The Helium flow measured on point
1 or 2 cannot be adjusted
Flow sensor 1 or 2.
Replace flow sensor
EVP1 or EVP2
Check voltage supply
Replace the solenoid valve
The pneumatic circuit is perfectly
close but the flow value do not
decrease up to zero performing the
Leak Test.
EV3 and/or EV4
Check voltage supply
Replace the solenoid valve