2
Maintaining the DSQ II
Maintaining the Rotary-Vane Pump
66
DSQ II Hardware Manual
Thermo Scientific
Figure 38.
Rotary Vane Pump Components
a.
Set the gas-ballast control to position I (means low
flow).
b. Leave it in that position for 10 minutes or until the oil
is clear.
c.
If the oil remains cloudy or discolored after 10
minutes, change the oil, as described in "
Changing the Oil in a Rotary-Vane Pump"
.
d. Set the gas-ballast control to position (0 means closed).
1
Foreline Vacuum Hose
8
Drain Plug
2
Inlet Port
9
MIN (minimum level mark)
3
Gas-Ballast Control
10
Mode Selector
4
Oil Filler Plugs
11
On/Off Switch
5
Outlet Port
12
Handle
6
MAX (maximum level mark)
13
Voltage Indicator
7
Oil-Level Sight-glass
14
Electrical Inlet Connector
2
14
3
1
5
6
10
9
11
12
7
4
8
13
Closed
Low
High
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...