Maintaining the DSQ II
Replacing the Electron Multiplier
60
DSQ II Hardware Manual
Thermo Scientific
5. Remove the ion detector.
Figure 34.
Removing the Ion Detector
a.
Disconnect the conversion dynode and electron multiplier connectors from the
connectors on the bottom of the vacuum manifold.
b. Remove the two screws holding the ion detector to the vacuum manifold.
6. Disassemble the ion detector and remove the multiplier.
1
Ion Detector
4
Two Screws on Ion Detector
2
Conversion Dynode Connector
5
Vacuum Manifold
3
Electron Multiplier Connector
1
4
2
3
5