4
Vacuum System and Gas Inlets
Vacuum System
156
DSQ II Hardware Manual
Thermo Scientific
High Vacuum Pump (Single Turbomolecular Pump)
The high vacuum pump creates the low pressure require by the ion source, quadrupole, and
ion detector. The DSQ II
can be equipped with one of two types of single turbomolecular
pumps: a 70 L/s or a 250 L/s.
In this configuration, the ion source, quadrupole, and ion detector are pumped by a single
pump. The 250 L/s pump is recommended for applications that use high gas flows into the
vacuum manifold, such as CI.
Turbomolecular pump operation is purely mechanical and does not require oil to create a high
vacuum. These turbomolecular pumps have a multi-stage axial-flow turbine in which
high-speed rotating blades provide compression by increasing the probability of gas molecules
moving in the pumping direction. The turbomolecular pump requires the use of a rotary-vane
pump to exhaust to atmosphere.
Figure 76.
70 L/s Turbomolecular Pump
1
Turbomolecular Pump Outlet
2
Electrical Supply Connector
1
2
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...