3
Troubleshooting
Heated Zone Symptoms
Thermo Scientific
DSQ II Hardware Manual
133
Ion source heater will not heat.
Possible Solutions
• The 24 V power supply is not operable. If necessary, replace power module, as described
in
• There is a problem with the analog board. Contact Technical Support.
• Cartridge heaters are not connected. Check connections between the analog and lens
interface boards and lens interface board and ion source board.
• One or more heater cartridges on the ion source board is defective. Vent the system and
check that the resistance of the heater circuit
≅
24
Ω
. If necessary, replace the ion source
board, as described in
“Cleaning the Ion Source, Lenses and Prefilter”
.
• Ion source heater fuse (F1) is blown on the analog board. Replace the ion source heater
fuse on the analog board, as described in
“Replacing the Analog Board Fuse”
• Ion source temperature sensor (RTD) is defective. Vent the system and check the
resistance of the RTD (108
Ω
at 20 °C, 176
Ω
at 200 °C). If necessary, replace the ion
source board, as described in
“Cleaning the Ion Source, Lenses and Prefilter”
.
Ion source heater overheats.
Possible Solutions
• The ion source temperature sensor (RTD) is defective. Vent the system and check the
resistance of the RTD (108
Ω
at 20 °C, 176
Ω
at 200 °C). Replace the ion source board if
necessary.
• There is a problem with the analog board. If necessary, replace the analog board, as
described in
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...