2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
94
DSQ II Hardware Manual
Thermo Scientific
2. Remove and replace the inlet valve seal.
Figure 53.
Removing the Inlet Valve Seal
a.
Insert the inlet seal removal tool into the inlet valve.
b. Press the button on the tool to engage the inlet valve seal.
c.
Pull out the inlet valve seal and discard it.
d. Insert a new inlet valve seal (PN 119265-0003).
3. Reassemble the inlet valve.
a.
Put the ferrule, knob and plug back into the inlet valve opening.
b. The first time you use the inlet valve, you will need to tighten the inlet valve knob
more than usual to prevent it from leaking.
Replacing the Ion Gauge (Optional Upgrade)
To replace the ion gauge, you will need:
• Gloves, clean, lint- and powder-free
• Ion gauge (PN A0105-06003)
• Wrench, Allen, 3 mm
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
Inlet Seal Removal Tools
CAUTION - INSTRUMENT DAMAGE.
To avoid scratching the sealing surface, you should
only use the supplied inlet seal removal tool to remove the seal.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...