Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
86
DSQ II Hardware Manual
Thermo Scientific
a.
Remove the lower screw and only loosen the upper screw holding the DC driver
cover in place.
b. Slide the cover up off the upper screw to remove it.
c.
Remove the DC driver cover.
5. Remove the current DC driver fuse.
6. Replace the DC driver fuse (PN 76703-0125; 1.25 A, SB, 5 x 20 mm, 250 V).
Figure 47.
Replacing the DC Driver Fuse
7. Reattach the DC driver cover and reattach the screw that holds the cover in place.
8. Reattach the left, top, and front covers of the DSQ II.
9. Restart the DSQ II.
DC Driver Fuse
CAUTION - INSTRUMENT DAMAGE.
Only use a 125V fuse or else you could damage the
instrument. Never replace DSQ II fuses with those of a different type, voltage, or current
rating. Also, to avoid damaging the boards, you should always wear an electrostatic
discharge (ESD) strap on your wrist when replacing fuses.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...