UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Page
78
Operating Instructions
4.6 Column Compartment
4.6.1 Installing a Column Switching Valve
The following column switching valves are available:
Column Switching Valve
Part No.
2-position, 6-port switching valve
6041.0004
2-position, 10-port switching valve
6041.0001
1.
Do not touch any metal or plastic parts inside the column chamber while the
temperature is higher than 50 °C. Before carrying out any work in the column
compartment, wait for the column chamber to cool down.
To shorten the cool-down time, consider setting a lower temperature and opening the
front panel door.
2.
A lever is provided on the enclosure bottom of the column compartment for pulling
forward the valve actuators. Turn the lever to right and pull it toward the front as far as
it goes out.
Fig. 35: Pulling the valve actuator forward
Tip:
Closing the front panel door will move the lever, and thus the actuators
automatically backward.
3.
Align the two pins on the back of the valve with the matching openings on the
actuator.
Fig. 36: Aligning the valve
Содержание Dionex UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Operating Instructions ...
Страница 8: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 16 Operating Instructions ...
Страница 62: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 54 Operating Instructions ...
Страница 92: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 84 Operating Instructions ...
Страница 246: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 238 Operating Instructions ...
Страница 262: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 254 Operating Instructions ...