![Thermo Scientific Dionex UltiMate 3000 Series Скачать руководство пользователя страница 215](http://html.mh-extra.com/html/thermo-scientific/dionex-ultimate-3000-series/dionex-ultimate-3000-series_operating-instructions-manual_1105017215.webp)
UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Operating Instructions
Page
207
6.
Connect the other end of the elbow tubing to the pump head inlet.
Fig. 80: Installing a shut off valve in a solvent line
Removing Solvent Shut Off Valves
1.
Stop the pump flow and wait until the pressure is down to zero.
2.
Close the shut off valve.
3.
Disconnect the elbow tubing from the pump head and empty the line into waste.
4.
Unscrew the cap of the solvent reservoir and remove the solvent line together with the
cap from the reservoir.
5.
Empty the solvent line to the reservoir into waste.
6.
Disconnect the shut off valve with the elbow tubing from the solvent line to the
reservoir.
7.
Connect the solvent line between pump head inlet and solvent reservoir.
No.
Description
1
Elbow tubing to pump head
2
Shut off valve
3
Solvent line to solvent reservoir
Содержание Dionex UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Operating Instructions ...
Страница 8: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 16 Operating Instructions ...
Страница 62: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 54 Operating Instructions ...
Страница 92: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 84 Operating Instructions ...
Страница 246: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 238 Operating Instructions ...
Страница 262: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 254 Operating Instructions ...