Dionex ICS-5000
+
Installation Instructions
40
Doc. 065447-02 12/14
3.6.2
Installing the Columns in the TC
The TC can hold up to three separator columns with inner diameters of 1 to
9 mm and maximum lengths of 30 cm (11.8 in).
1. This section provides brief installation instructions for the guard column
and separator column. Refer to the column manual for detailed
installation and start-up instructions. Column manuals are provided on the
Thermo Scientific Reference Library DVD (P/N 053891).
2. Remove the guard column and separator column from their boxes.
3. Remove the fitting plugs from the ends of each column.
4. The column clip (P/N 064786) consists of two pieces: a base and a
retaining ring (see
5. Remove the retaining ring from the clip by pressing lightly on the ring in
the direction indicated by the arrows on the ring.
6. Attach the base of the clip to a column bracket,
and then rotate the clip 90 degrees clockwise (see
7. Press the column into the clip and reinstall the
ring.
Figure 3-12. Column Clips Installed
Figure 3-11. TC Column Clips
Base
Retaining
Ring
Arrows