4 • Troubleshooting
Doc. 031857-03 4/06
4-13
•
Damaged pump piston seal
1. Replace the piston seal (see
2. If the leak persists, replace the piston (see
•
Pump head not tight against casting
Carefully tighten the pump head mounting nuts just until the leak stops.
DO
NOT OVERTIGHTEN!
•
Leaking pressure transducer
1. Make sure the liquid line connections into the pressure transducer are
tight. Refer to
Installation of Dionex Liquid Line Fittings
(Document No.
031432) for tightening requirements. The manual is on the Dionex
Reference Library CD-ROM (P/N 053891). Replace any damaged
fittings.
2. If the pressure transducer continues to leak, contact Dionex for
assistance.The pressure transducer cannot be serviced by the user.
•
Leaking pump head waste valve
Make sure the waste valve is closed. To close the valve, turn the knob
clockwise, and tighten fingertight.
DO NOT OVERTIGHTEN!
Overtightening may damage the valve and the pump head.
If the leak persists, replace the waste valve O-ring (see
•
Leaking suppressor
Refer to the suppressor manual for troubleshooting procedures. Suppressor
manuals are included on the Dionex Reference Library CD-ROM
(P/N 053891).
•
Leaking injection valve
1. Make sure the liquid line connections to the transducer are tight. Refer to
Installation of Dionex Liquid Line Fittings
(Document No. 031432) for
tightening requirements. The manual is on the Dionex Reference Library
CD-ROM (P/N 053891). Replace any damaged fittings.
2. Liquid leaks from behind the valve stator may indicate a scratched rotor
seal. Rebuild the injection valve (see
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...