Setting Up the Eluent Generator
Doc. 031857-03 4/06
B-87
7. Direct the vent line to the waste container and insert it down into the
gas separator waste tube about 10 to 15 cm (4 to 6 in).
Recording the EluGen Cartridge Serial Number in Chromeleon
The serial number is printed on the EluGen cartridge label.
1. Open the Chromeleon Server Configuration program (select
Start>Programs
(or
All Programs
)
>Chromeleon>Server
Configuration
).
2. Double-click the ICS-2000 system icon. The properties dialog box for
the ICS-2000 appears.
3. Click the
Eluent Generator
4. Select the
EGC-1
check box.
5. Enter the EluGen cartridge serial number and click
OK
.
Figure B-86. Eluent Generator Configuration Properties Example