ICS-2000 Ion Chromatography System
4-14
Doc. 031857-03 4/06
•
Leaking cell
1. Check the waste lines for blockage; trapped particles can plug the lines
and cause a restriction and/or leak. If necessary, clear the waste lines by
reversing the direction of flow.
2. Make sure the plumbing downstream from the cell is clear; a blockage
may overpressurize the cell, causing it to leak. If the problem continues,
contact Dionex for assistance.
4.3
Pump Difficult to Prime or Loses Prime
Excessive pressure fluctuations (more than 3% difference from one pressure
reading to the next) indicate that the pump is out of prime.
•
Empty eluent reservoir and/or no eluent connected
Fill the reservoir. Make sure all connections are secure.
•
Eluent improperly or insufficiently degassed
Check the vacuum degas settings:
1. Open the Chromeleon Server Configuration program, select the ICS-2000
device icon under the timebase, and select
Edit>Properties
(or double-
click the ICS-2000 device icon).