2 • Features
Doc. 065289-03 10/12
27
Select the
Options
.
Figure 2-10. Server Configuration Properties: Degas Options
Degas Options
•
Always Off
: The degas pump is always off.
•
Always On
: The degas pump is always on. This setting is for test
purposes by a Thermo Fisher Scientific Service Representative. Do
not use this setting for routine operation.
•
Cycle
: The degas pump cycles on and off according to the times
specified in the
Cycle On
and
Off
fields.
Cycle On
specifies for how
long the degas pump runs during a cycle.
Cycle Off
specifies the time
between cycles.
•
Monitor
: (default mode) The Dionex ICS-1100 monitors the degas
pressure reading and turns the degas pump on and off as required.
Содержание Dionex ICS-1100