Doc. 065289-03 10/12
The outputs can be used to control functions in external devices such as an
autosampler or another Dionex module. When connected to a controlling device,
the inputs can be programmed to perform the following Dionex ICS-1100
functions:
•
Switch the injection valve position (load/inject)
•
Perform an autozero command (set the conductivity to zero)
•
Turn the pump on and off (also turns the suppressor on and off)
•
Send a chart mark signal to the analog output. The mark is 10% of the full-
scale voltage, and the duration is 0.5 seconds. A mark can be used, for
example, to indicate the injection.
Relay outputs 1 and 2 can be programmed to switch any low-voltage control.
Switched current must be less than 200 mA and 42 V peak.
Connecting a TTL or Relay
1. Locate the twisted pair of
wires (P/N 043598) and the
12-position connector plug
(P/N 923687) (see
) in the Dionex
ICS-1100 Ship Kit (P/N
057905).
2. Follow these basic steps to
connect the TTL or relays.
a.
For each relay or TTL to
be used, connect an active wire (red) and a ground wire (black) to the 12-
position connector plug at the appropriate pin locations. Refer to
or the label on the Dionex ICS-1100 rear panel for the
connector pin assignments.
To attach a wire to the plug, strip the end of the wire, insert it into the
plug, and use a screwdriver to tighten the locking screw. If necessary,
multiple ground wires can be attached to a single TTL input/output
ground pin.
When attaching wires to the connector plug, be careful not to allow
stray strands of wire to short to the adjoining position on the
connector.
Figure B-2. 12-Position Connector Plug
Position 1
Position 12
Locking
Screws
Содержание Dionex ICS-1100
Страница 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Страница 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Страница 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Страница 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Страница 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Страница 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Страница 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Страница 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...