11. Align the top of the suppressor
with the guide line printed on the
component panel (see
) and align the slots
on the rear of the suppressor
with the tabs on the panel.
12. Press the suppressor onto the
tabs and then slide it down until
it locks into place. Pull slightly
on the center of the suppressor to
verify that it is securely fastened.
Note
: Some suppressors require
more force to secure them onto
the tabs.
13. Connect the two eluent and two
regenerant lines to the new
suppressor.
14. Close the Dionex ICS-1100 front
door.
Figure 5-15. Suppressor Guidelines and Mounting
Tabs on Component Panel
5.12 Replacing the Column Heater
1. Turn off the Dionex ICS-1100 power.
2. Open the front door of the Dionex ICS-1100.
3. Remove the existing column heater:
a.
Unscrew the two thumbscrews on the heater cover (they remain attached
to the cover).
b. Pull the heater cover straight out to remove it.
c.
Unscrew the six thumbscrews on the top metal plate (see
remove the plate.
1
1
2
3
4
1
Suppressor
Mounting Tabs
Suppressor
Cable Connector
Guideline for
AES
Guideline for
SRS
1
2
3
4
Содержание Dionex ICS-1100
Страница 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Страница 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Страница 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Страница 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Страница 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Страница 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Страница 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Страница 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...