Dionex ICS-1100 Ion Chromatography System
92
Doc. 065289-03 10/12
5.5
Replacing an Auxiliary Valve Pod
This procedure describes how to replace the mechanical part (the “pod”) of the
optional auxiliary valve. This procedure is an alternative to rebuilding the valve
(see
NOTE If the valve electronics require service, contact Thermo
Fisher Scientific. Electronics-related repair procedures
must be performed by Thermo Fisher Scientific
personnel.
1. Turn off the pump flow from the Dionex ICS-1100 Control panel in
Chromeleon.
2. Open the Dionex ICS-1100 front door.
3. Disconnect each liquid line connected to the valve.
Figure 5-4. Injection Valve Plumbing
Substituting non-Dionex/Thermo Fisher Scientific parts may impair
performance, thereby voiding the product warranty. Refer to the
warranty statement in the Dionex Terms and Conditions for more
information.