To reinstall the pump head and piston:
Thermo Fisher Scientific recommends reinstalling the pump head and piston as a
single assembly, so that the piston centers itself onto the magnetic follower.
1. Hold the assembled spacer and guide with the drain tube aligned vertically,
and then press the spacer into the pump head until it is flush with the indented
surface of the head.
2. Insert the piston so that about 6 mm
(1/4 in) of the shaft remains exposed. This
ensures that the magnet in the follower
picks up the piston. (The follower is the
cylinder that holds the piston in place as it
moves in and out of the pump head
assembly.)
3. Reinstall the pump head and piston
assembly. Tighten the screws evenly.
To complete the procedure:
1. Reconnect all liquid lines to the pump head.
2. Close the waste valve knob.
3. Turn on the Dionex Easion power switch.
5.9
Replacing a Pump Piston
Continued leaking of the piston seal after installation of a new seal (assuming the
pump head is tight) indicates a dirty, scratched, or broken piston.
Follow the instructions in
to install a new piston (P/N 052840) and
piston seal (P/N 055870). Always replace the piston seal when replacing a piston.
Содержание Dionex Easion
Страница 1: ...Dionex Easion Ion Chromatography System Operator s Manual Document No 155028 Revision 01 August 2020 ...
Страница 34: ...Dionex Easion Operator s Manual 26 Doc 155028 01 08 20 ...
Страница 68: ...Dionex Easion Operator s Manual 60 Doc 155028 01 08 20 ...
Страница 76: ...Dionex Easion Operator s Manual 68 Doc 155028 01 08 20 Figure 5 2 Dionex Easion Flow Schematic ...
Страница 110: ...Dionex Easion Operator s Manual 102 Doc 155028 01 08 20 ...
Страница 118: ...Dionex Easion Operator s Manual 110 Doc 155028 01 08 20 ...