5 • Service
Doc. 065259-05 10/12
81
7. Reinstall the valve assembly and reconnect all liquid lines.
8. Turn on the power to the Dionex AS-DV.
9. Start a test run to check for leaks from the valve. Tighten fittings as required.
10. Reinstall the front cover and close the top cover.
5.6
Replacing a High-Pressure Valve Pod
This procedure describes how to replace the mechanical parts (the “pod”) of an
optional high-pressure valve.
NOTE Substitution of non-Dionex/Thermo Scientific parts may
impair valve performance and void the product
warranty.
The following tool is needed to complete this procedure: 3-mm Allen wrench.
1. Turn off the Dionex AS-DV power.
2. Raise the top cover.
3. Remove the Dionex AS-DV front cover by gripping the indentations at each
side and pulling the cover straight off toward you.
4. Disconnect each liquid line connected to the valve.
The Dionex AS-DV power must be turned off before you replace a
valve pod. If the power is on when you disconnect or connect the
valve cable, the Dionex AS-DV may be damaged.
Содержание Dionex AS-DV