Diagnostics
Device Status
Thermo Fisher Scientific
DensityPRO with
F
OUNDATION
™ fieldbus
Application Guide
8-3
5.
Does set output x mA equal readout device?:
Confirm the set output
and reference meter reading are the same. Enter
1
to continue to the
next step. If you need to modify the value again, enter
2
.
6.
If you entered
1
in the previous step, the software will show that the
current output is being set to maximum.
7.
Enter meter value:
Enter the value displayed by the reference meter.
8.
Does set output x mA equal readout device?:
Confirm the set output
and reference meter reading are the same. Enter
1
to continue to the
next step. If you need to modify the value again, enter
2
.
9.
If you entered
1
in the previous step, the software will show that the
field device is being returned to its original output.
10.
Click
Next
to finish the process. The loop may be returned to
automatic control.
Device Status
When you select the Device Status item from the Diag/Service menu, any
alarms or bad flags (indicators that the gauge has been incorrectly
configured) are displayed. Click
Next
to view each one. An example is
shown below.
Figure 8–1.
Device status example
Содержание DensityPRO
Страница 2: ......
Страница 3: ...DensityPRO With FOUNDATIONTM Fieldbus Application Guide P N 717917 Revision A...
Страница 4: ......
Страница 6: ...This page intentionally left blank...
Страница 8: ...This page intentionally left blank...
Страница 44: ...This page intentionally left blank...
Страница 48: ...This page intentionally left blank...
Страница 52: ...This page intentionally left blank...
Страница 56: ...This page intentionally left blank...
Страница 62: ...This page intentionally left blank...
Страница 78: ...This page intentionally left blank...
Страница 94: ...This page intentionally left blank...
Страница 96: ...This page intentionally left blank...