6-2
DB-070 E
RadEye AB100
Pi/Ff 04.05.2021
6.2
Failure indication
In case of a failure the beeper generates a melody every 32 s.
The corresponding failure message is displayed in the LCD:
Error high voltage generation
No detector pulse within configured
time
EEPROM with calibration data shows
EEPROM Read or EEPROM Write
error.
A Watchdog Error indicates that the
microcontroller has problems to work
on its tasks in a given timeframe.
Reasons are strong electromagnetic
pulses, firmware errors or hardware
issues.