![Thermo Scientific Cryofuge 16 Скачать руководство пользователя страница 64](http://html.mh-extra.com/html/thermo-scientific/cryofuge-16/cryofuge-16_instruction-manual_1104994064.webp)
B–7
Thermo Scientific Cryofuge 8 / 16 Centrifuges and Heavy Duty Centrifuges
Rotors
HAEMAFlex 8 Rotor
with Single Blood Bag Buckets with Filter Pack
Technical Data
Cryofuge 8
Centrifuge Voltage
200, 208, 220,
230, 240 V
380, 400, 415 V
Rotor Capacity
8 x 1000 ml
8 x 1000 ml
Maximum Speed n
max
4
600 rpm
4
600 rpm
Maximum RCF-Value at n
max
7
073 x g
7
073 x g
K-Factor at n
max
7
942
7
942
Radius max. / min.
299 mm / 154 mm
299 mm / 154 mm
Angle
90°
90°
Maximum Speed at 4 °C
4
600 rpm
4
600 rpm
Sample Temperature
at Max. Speed
(Ambient temperature of 23 °C,
run time 90 minutes)
50 Hz: 4 °C
60 Hz: 2 °C
50 Hz: 4 °C
60 Hz: 2 °C
Aerosol-tight
No
No
Rotor Body
Max. Cycle Number
50 000
Buckets
Max. Cycle Number
1
at
7073 x g / 4600 rpm
5000 x g / 3868 rpm
4000 x g / 3459 rpm
2500 x g / 2735 rpm
14 000
31 000
50 000
138 000
Weight
28.8 kg
Rotor Body (empty)
Windshield Lid
Windshield Bowl
21.9 kg
1.9 kg
5.0 kg
1
Cycle numbers stated are only valid, if the according speed (or rcf) is always set for centrifugation. If different speed (or rcf) settings are used, the
smallest cycle number stated is valid.
Table B–7: