5. Connect the in-line filter to the system flow path in a post-column
position. Observe the direction of flow as indicated on the filter.
a) Connect the capillary from the column outlet to the upstream
end of the in-line filter and route it to the waste reservoir.
b) Flush the in-line filter with mobile phase at a flow rate of about
5 mL/min.
c) Connect the downstream end of the in-line filter as required by
the system flow path (see next section).
5.6.5
Connecting the Detector in the System Flow Path
Depending on the modules in your HPLC system, you can connect the
inlet capillary to the detector as follows:
•
Directly from the column
–or–
•
From the flow cell of a first detector, such as a diode array detector,
in the HPLC system flow path (series flow connection)
In a series liquid flow configuration, one flow path from the pump is
used for both detectors installed in the system.
TIP
Alternatively, you can connect the detector from a flow-splitting
device (parallel flow connection). In a parallel liquid flow configuration,
one flow path from the pump is split for two detecting devices. For
further information on series and parallel liquid flow configurations, see
Parts required
From the detector ship kit:
•
RheFlex fittings
•
Inlet capillary, PEEK, 0.005" x 1/16" (I.D. x O.D.), red
TIP
Keep the capillary connection to the charged aerosol detector as
short as possible to minimize peak dispersion.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...