Cell Locker
50155007 Revision C May 2021
Operating Instructions
Страница 1: ...Cell Locker 50155007_C 27 05 21 Thermo Fisher Scientific Cell Locker 50155007 Revision C May 2021 Operating Instructions ...
Страница 2: ...ssions damage or loss that might result from any use of this document even if the information in the document is followed properly This document is not part of any sales contract between Thermo Fisher Scientific Inc and a purchaser This document shall in no way govern or modify any Terms and Conditions In case of conflicting translations into foreign languages the German language version of these ...
Страница 3: ...correct Use 4 Chapter 2 Delivery of the device 5 2 1 Scope of delivery 5 2 2 Acceptance inspection 5 Chapter 3 Description of the device 6 3 1 Description 6 3 2 First use 7 3 3 Operation 7 3 4 Application versions 7 Chapter 4 Cleaning and disinfection 9 4 1 Cleaning 9 4 2 Wipe spray disinfection 10 4 3 Autoclaving 11 Chapter 5 Service 12 5 1 Membrane filter replacement 12 5 2 Cover gasket replacem...
Страница 4: ...erate injuries Notice Indicates a situation which if not avoided could result in damage to equipment or property In chapters of the instruction manual which deal with safety this icon appears under the title of the chapter Displayed on the equipment this icon denotes that special attention must be paid to the information given in the instruction manual or accompanying documents Is used for applica...
Страница 5: ...the sample weight limits as a whole and its shelving in particular Check the door seal every six months for proper sealing performance and possible damage Do not process any samples containing hazardous chemical substances that may be released into the ambient air through defective seals or may cause corrosion or other defects 1 4 Warranty Thermo Fisher Scientific warrants the operational safety a...
Страница 6: ...maximum Autoclave temperature is 15 minutes at 121 C 250 F with fast exhaust 1 5 2 Incorrect Use Do not use tissues substances or liquids that are easily flammable or explosive release vapor or dust that forms combustible or explosive mixtures when exposed to air release poisons release dust exhibit exothermic reactions are pyrotechnical substances Do not pour any liquids on the bottom of the inte...
Страница 7: ...OCKER has been delivered check the delivery immediately for completeness possible damage If components are missing or damage is found on the device or the packaging please contact the carrier as well as the customer service immediately Description Pcs CELL LOCKER 1 Membrane filter 2 Work tray with cover 1 Consumables kit 1 Operating instructions 1 Summarized safety instructions 1 ...
Страница 8: ...CKER consists of a polycarbonate chamber with two membrane filters which are permeable to water vapor and gas but are impermeable to microorganisms Fig 1 CELL LOCKER Fig 2 Bottom marks Position Name 1 work plate 2 membrane frame with filter 3 gasket 4 latch 5 cover 6 handle knob 1 2 3 4 5 6 Production year Production month Polycarbonate Makrolon 2528 ...
Страница 9: ... cover When inserting into a laboratory device make sure that the CELL LOCKER is not twisted 3 4 Application versions Depending on user preference the CELL LOCKER can be modified to attach or remove the transport cover using the consumables kit Fig 3 Consumables kit Note Clean the CELL LOCKER before inserting the membrane frame with filter Warning The CELL LOCKER must be removed before the incubat...
Страница 10: ... CELL LOCKER options with attached transport cover without transport cover without cover to remove cultures with single door access with cover to remove CELL LOCKER with cultures samples access inside the chamber with transport cover Note Attach cover for protected transport to biosafety cabinet or to another incubator or lab Use of cover in the chamber may require two hand opening ...
Страница 11: ... cleaning liquids may cause infections Disinfectants may contain harmful substances When cleaning and disinfecting always observe the safety instructions and hygiene guidelines Wear safety goggles Wear safety gloves Wear breathing protection Observe the safety instructions of the disinfectant s manufacturer and the hygiene supervisor Caution Incompatible cleaners Some of the CELL LOCKER components...
Страница 12: ...aces air baffle and internals dry with a soft sterile cloth Caution Alcoholic disinfectants Disinfectants having a alcohol content of more than 10 may form in combination with air easily combustible and explosive gas mixtures When using such disinfectants avoid open flames or exposure to excessive heat during the entire disinfection process Use such disinfectants only in adequately ventilated room...
Страница 13: ... The CELL LOCKER can be autoclaved up to 15 minutes at 121 C with fast exhaust Autoclaving at higher temperature is not recommended since the material cannot withstand higher temperature and its composition would be negatively affected Note Remove the membrane frame with filter before autoclaving the CELL LOCKER ...
Страница 14: ...r replacement Periodically check the membrane frame filters and at least once a year replace them 1 Pull out the work tray and take it out 2 Press and remove the membrane frame from the inside 3 Put the membranes into a plastic bag autoclave them and dispose as hazardous waste Warning Health hazard Wear safety gloves Wear breathing protection ...
Страница 15: ...adial direction 6 Visually inspect from the inside whether the gasket lip is correctly seated if necessary press from the outside 7 Insert the second membrane according to 5 and 6 8 Insert the work tray over the tilting stop and push into the interior Warning Contamination hazard Damaged or improperly mounted membranes can contaminate the user or the environment ...
Страница 16: ...integrity no damage and replace as needed 1 Pull out the work tray and take it out 2 Remove cover gasket from the notch 3 Insert a new gasket in the notch and take care that the gasket lip is outside 4 Insert the work tray and push into the interior Warning Contamination hazard Damaged or improperly mounted membranes can contaminate the user or the environment ...
Страница 17: ...ccessories that have been approved by Thermo Fisher Scientific used third party spares without Thermo Fisher Scientific approval void the limited warranty Description Material No Membrane frame s 50153148 Cover seal 6 pieces 50156433 Work tray stainless steel 50155808 Work tray copper 50155809 Consumables kit 50161834 Cover CELL LOCKER 50156434 ...
Страница 18: ...processing infectious substances Therefore the device and device components may have been contaminated Prior to disposal all device components must be decontaminated Clean the unit components thoroughly then disinfect decontaminate or autoclave them depending on application Attach a declaration of non objection with details of performed decontamination measures to the items that are to be disposed...
Страница 19: ...ical Data 8 Technical Data Description Unit Value Outer dimensions W x H x D mm 185 x 156 x 380 Inner space volume l approx 5 4 Weight kg 1 8 Loading capacity per work tray kg 5 Operating temperature C F 4 55 39 131 Autoclave cycles max 12 Autoclave temperature C F max 121 250 ...
Страница 20: ......